MPI-520 – USER MANUAL
Test range according to IEC 61557-2 for U
N
= 1000 V: 1000 k…3,00 G
Display range for
U
N
= 1000 V
Test voltages: 50 V, 100 V, 250 V, 500 V and 1000 V
Accuracy of generated voltage (Robc [] 1000*U
N
[V]): -0+10% from the set value
Detection of a dangerous voltage before commencing a measurement
Discharging the object tested
Measurement of insulation resistance with the use of UNI-Schuko plug (WS-03, WS-04) be-
tween all three terminals (U
N
=1000 V is not available)
Insulation resistance measurement for multi-wire cables (max. 5) using an optional external
adapter
Measurement of voltage on terminals +R
ISO
, -R
ISO
within the range of: 0..440 V
Test current < 2 mA
Remark:
With regard to measurements conducted with the use of WS-03 and WS-04 leads, if at
least one out of three measurements has ended with current limitation (LIMIT symbol
is displayed), the results of remaining measurements may be burdened with additional
uncertainty.
Phase sequence
Phase sequence indicator: correct, incorrect
Mains voltage range U
L-L
: 95...500 V (45…65 Hz)
Display of phase-to-phase voltages
Other technical data
a) type of insulation ..................................................... double, EN 61010-1 and IEC 61557 compliant
b) metrological category..................................................... IV 300V (III 600V), EN 61010-1 compliant
c) degree of housing protection acc. to EN 60529........................................................................ IP54
d) power supply of the meter...............................................................................................................
............... alkaline batteries 4x1,5 V LR14 (C) or accumulator package SONEL NiMH 4,8 V 4,2 Ah
e) parameters of AC adapter for the battery charge .............................. 100 V…240 V, 50 Hz…60 Hz
f) dimensions ....................................................................................................... 288 x 223 x 75 mm
g) weight of the meter ......................................................................................................about 2,2 kg
h) storage temperature .................................................................................................... –20...+70C
i) working temperature ........................................................................................................ 0...+50C
j) temperature range suitable for initiating battery charging ....................................... +10C to +40C
k) temperatures at which loading is interrupted ................. below +5C and above (or equal to) +50C
l) humidity ......................................................................................................................... 20%...90%
m) nominal temperature ....................................................................................................... +23 ± 2C
n) reference humidity ......................................................................................................... 40%...60%
o) altitude (above sea level) ................................................................................................... <2000m
p) time till automatic shutdown when idle (Auto-OFF) ..................................................... 120 seconds
q) number of measurements Z or RCD (for alkaline batteries) ... >3000 (2 measurements per minute)
r) number of measurements R
ISO
or R (for alkaline batteries) ................................................... >2000
s) display .............................................................................................................. LCD, segment-type