29- For verification purposes, the date is required to be entered for a second time. Enter the
calibration date (year, month & date) then click Continue
If the Continue button is pressed without entering a date, close the SphygmoCor XCEL Calibration
software, turn OFF and then ON the SphygmoCor XCEL device, and repeat the calibration
procedures from step 1. Please refer to Restarting the Calibration Procedure Section in the
Calibration Troubleshooting section.
Note: If this message appears (incorrect date entered), enter the correct date and press Continue.