SphygmoCor XCEL System

Questions and Answers

  • S
    Sara DodsonSep 12, 2025
    What to do if SphygmoCor XCEL is unable to continue Brachial Calibration due to weak oscillometric signal?
    • M
      michelleyatesSep 12, 2025
      If you're unable to continue Brachial Calibration due to a weak oscillometric signal on your SphygmoCor XCEL Medical Equipment, ensure the correct cuff size is being used, with correct positioning and fit. Also, make sure there are no obstructions, such as clothing, between the cuff and the patient’s arm. After these steps, attempt another measurement.
  • F
    fhensonSep 9, 2025
    What to do if Tonometer fails to respond on SphygmoCor XCEL?
    • M
      Michael JohnsonSep 10, 2025
      If the Tonometer fails to respond with the SphygmoCor XCEL device, ensure that the Tonometer is connected to the appropriate port at the rear of the SphygmoCor XCEL device.
  • B
    brooksryanSep 8, 2025
    What to do if SphygmoCor XCEL is unable to generate PWV Report?
    • H
      holmesjasonSep 8, 2025
      If you are unable to generate a PWV Report on your SphygmoCor XCEL Medical Equipment, ensure that the entered distance in mm is correct. Verify the correct cuff size is used and that it is correctly positioned. Ensure waveform quality by using the guidance bar. After these steps, repeat the PWV measurement.
  • D
    Dennis EllisSep 7, 2025
    What to do if SphygmoCor XCEL Medical Equipment is unable to continue Capture due to deflation of cuff?
    • T
      tracy14Sep 7, 2025
      If your SphygmoCor XCEL Medical Equipment is unable to continue Capture due to deflation of cuff, ensure that the patient’s limb with the cuff is stationary. Use the correct cuff size and ensure it is correctly positioned. Ensure that there are no obstructions between the cuff and arm. After completing these steps, attempt Capture again.
  • B
    Benjamin BenderAug 23, 2025
    How to fix SphygmoCor Medical Equipment XCEL device if it cannot be detected?
    • W
      walkerricardoAug 23, 2025
      To resolve the issue where the SphygmoCor XCEL device cannot be detected, verify these items: Ensure the device is connected to your computer and plugged into an AC outlet with the power switch in the 'on' position. Check that the Power light is illuminated. If it's not, refer to condition 2 'The POWER light is off'. Confirm the correct communications port is selected in the configuration settings by selecting System – Find Module from the Setup Screen. If you're using a USB Adapter, ensure the USB drivers have been installed. If the problem persists, contact AtCor Medical Technical Support.
  • D
    Debbie ConradSep 5, 2025
    What to do if SphygmoCor XCEL is unable to continue Capture due to invalid data from the Electronics Module?
    • M
      Michael SanchezSep 5, 2025
      If you are unable to continue Capture due to invalid data from the Electronics Module on your SphygmoCor XCEL Medical Equipment, press the STOP button on the device. Ensure that the device is connected to the computer. In the Setup screen, click on System and select Find Module. Attempt Capture again.
  • J
    Jasmin RodriguezSep 4, 2025
    What to do if SphygmoCor Medical Equipment is unable to continue Brachial Calibration due to cuff overpressure?
    • E
      ericyangSep 4, 2025
      If you are unable to continue Brachial Calibration due to cuff overpressure on your SphygmoCor XCEL Medical Equipment, ensure the correct cuff size is used, and that it's correctly positioned and fitted. Also, check that the hose is not bent or pinched and that the cuff and hose are not beneath the patient. After completing these steps, attempt another measurement.
  • C
    cindydiazSep 3, 2025
    What to do if SphygmoCor XCEL is unable to continue Brachial Calibration due to inflate timeout?
    • M
      Michael WoodSep 3, 2025
      If you are unable to continue Brachial Calibration due to an inflate timeout on your SphygmoCor XCEL Medical Equipment, ensure that the hose connections to the device and cuff are intact and tight. Ensure the correct cuff size is used with correct positioning and fit, and that the cuff is not leaking air. After completing these steps, attempt another measurement.
  • R
    Ryan MorrisSep 1, 2025
    What to do if SphygmoCor Medical Equipment is unable to continue Capture due to unsuccessful deflation of cuff?
    • E
      Eric KellerSep 2, 2025
      If your SphygmoCor XCEL Medical Equipment is unable to continue Capture due to unsuccessful deflation of the cuff, press the STOP button on the device to deflate the cuff. Make sure that the device is connected to the computer and that the cuff is properly connected to the device. Ensure the correct cuff size is used and correctly positioned. Turn the device OFF and ON, then try Capture again.
  • J
    Jeremy WalkerAug 31, 2025
    What to do if SphygmoCor XCEL Medical Equipment is unable to continue Brachial Calibration due to safety timeout?
    • I
      Ian OlsonSep 1, 2025
      If you're unable to continue Brachial Calibration due to safety timeout on your SphygmoCor XCEL Medical Equipment, turn the device OFF and then turn it ON again. Ensure the correct cuff size is used with correct positioning and fit. Also, ensure that the patient’s arm is stationary during the measurement. After completing these steps, attempt another measurement.

Summary

Troubleshooting SphygmoCor XCEL

Device and Software Issues

Addresses common problems with the XCEL device, its software, and error codes.

Database Management

Backup and Restore Operations

Procedures for backing up and restoring the system database to protect patient data.

Basic System Care

Handling and Environmental Care

Guidelines for device stability, environmental factors, and avoiding damage.

Cleaning and Disinfection

Procedures for cleaning the device, tonometer, and cuffs, including disinfection.

Cuff Specific Care

Detailed instructions for the maintenance of PWV Thigh and PWA Brachial cuffs.

Calibrating SphygmoCor XCEL

Calibration Kit and Setup

Information on the calibration kit, its components, and initial setup procedures.

Calibration Procedure Steps

Detailed, sequential steps for performing the calibration of the SphygmoCor XCEL device.

Calibration Software Installation

Computer Requirements

Specifies the hardware and software prerequisites for installing the calibration software.

Installation Guide

Step-by-step instructions for installing the SphygmoCor XCEL calibration software.

Calibration Troubleshooting

Finding the Com Port

Instructions on how to locate the correct COM port for device communication.

Restarting the Calibration Process

Steps to follow if the calibration process fails or requires restarting.

Spare Parts List

Technical Specifications

Physical, Environmental, and Computer Specs

Details on device dimensions, operating conditions, and computer requirements for software.

Classification and Standards