IM-P323-32
CTLS Issue 5
5
SX80 and SX90 Quickstart Guide
Cleaning
Isopropyl alcohol may be used to clean labels. Do not use water or
water based products. A mild soap solution may be used to clean
other exterior surfaces.
Electromagnetic compatibility
This instrument conforms with the essential protection requirements
of the EMC Directive 89/336/EEC, by the application of a Technical
Construction File. It satisfies the general requirements of the
industrial environment defined in EN 61326.
Caution - Charged capacitors
Before removing an instrument from its sleeve, disconnect the
supply and wait at least two minutes to allow capacitors to
discharge. Avoid touching the exposed electronics of an instrument
when withdrawing it from the sleeve.
Safety symbols used on the instrument have the following meaning:
Caution, refer to
throughout
accompanying
documents.
Equipment
protected by
double insulation.
Installation category and pollution degree
This unit has been designed to conform to BS EN 61010 installation
Category II and Pollution Degree 2, defined as follows: Installation
Category II (CAT II) – The rated impulse voltage for equipment on
nominal 230 V supply is 2500 V.
Pollution Degree 2 – Normally only non-conductive pollution occurs.
However, a temporary conductivity caused by condensation must
be expected.