PERFORMANCE TESTS
5-7
Capacitance Accuracy Table
capacitance Frequency Range SR715 tol. + C tol. SR720 tol. + C tol.
1.0n 1 kHz 0 .99780n - 1.0022n .99930n - 1.0007n
1.0n 10 kHz 1 .99680n - 1.0032n .99880n - 1.0012n
1.0n 100 kHz 2 ----------------- .99780n - 1.0022n
10.0n 100 Hz 0 9.9780n - 10.022n 9.9930n - 10.007n
10.0n 1 kHz 1 9.9780n - 10.022n 9.9930n - 10.007n
10.0n 10 kHz 2 9.9680n - 10.032n 9.9880n - 10.012n
10.0n 100 kHz 2 ----------------- 9.9780n - 10.022n
100n 100 Hz 1 99.780n - 100.22n 99.930n - 100.07n
100n 1 kHz 2 99.780n - 100.22n 99.930n - 100.07n
100n 10 kHz 2 99.680n - 100.32n 99.880n - 100.12n
100n 100 kHz 3 ----------------- 99.780n - 100.22n
1.000µ 100 Hz 2 .99780µ - 1.0022µ .99930µ - 1.0007µ
1.000µ 1 kHz 2 .99780µ - 1.0022µ .99930µ - 1.0007µ
1.000µ 10 kHz 3 .99680µ - 1.0032µ .99880µ - 1.0012µ
10.000µ 100 Hz 2 9.9780µ - 10.022µ 9.9930µ - 10.007µ
10.000µ 1 kHz 3 9.9780µ - 10.022µ 9.9930µ - 10.007µ
4. This completes the performance verification tests.