7. Checking the measuring tool
7.1 Accuracy check
To prevent measuring errors, the accuracy of the
measuring tool must be checked
at regular intervals;
for example, each time before beginning work, or
after a heavy impact or extreme fluctuations in temperature.
Step 1:
The unit of measurement ° degrees and SENS 0.00° must
be selected!
Place the unit with the lower measuring sole on as horizon-
tal a surface as possible (e.g. a table) with the display side
facing the user. Determine the measurement.
Step 2:
Turn the unit by 180° in the same position.
Step 3:
The rear of the unit is now facing the user.
Compare the new measurement with the measurement
determined in step 1.
In the case of deviations > 0.05°,
the unit must be calibrated again (-> Calibration).
Δ
≤ 0.05° = OK
Δ
> 0.05° calibrate
1
3
1
2
3
180°