Status Pro – T430 User Guide
Green (Y-)-LED
0,5Hz-Flash
Levelling: Y-axis is more than 1000 steps away from calibrated•
levelling value
Levelling: Temperature gradient still too high•
Levelling demand (after power on): Device in warming phase •
(1 min)
Levelling demand (after power off): Warning for unreliable •
values
2-Sensor-Control:•
Y-axis is more than 5 steps away from target position
Permanently lit
Levelling: Y-axis is levelled•
Calibration of Y-axis successful•
2-Sensor-Control:•
Y-axis completely arranged, target position reached
5Hz-Flash
Levelling: Y-axis is more than 1000 steps away from calibrated•
levelling value
Levelling: Y-(Z-)axis displays restive values•
Levelling not possible, outside measurement range•
2-Sensor-Control:•
Y-axis is more than 5 steps away from target position
0,5Hz-Flash
Levelling: Y-axis is more than 1000 steps away from calibrated•
levelling value
Levelling: Temperature gradient still too high•
Levelling demand (after power on): Device in warming phase•
(1 min)
Levelling demand (after power off): Warning for unreliable •
values
2-Sensor-Control:•
Y-axis is more than 5 steps away from target position