22
Status Pro – T430 User Guide
Green (X+)-LED
Green (X-)-LED
Permanently lit
Levelling: X-axis is levelled•
Calibration of X-axis successful•
2-Sensor-Control:•
X-axis completely arranged, target position reached
5Hz-Flash
Levelling: Y-axis is more than 1000 steps away from calibrated•
levelling value
Levelling: Y-(Z-)axis displays restive values•
Levelling not possible, outside measurement range•
2-Sensor-Control:•
Y-axis is more than 5 steps away from target position
0,5Hz-Flash
Levelling: Y-axis is more than 1000 steps away from calibrated•
levelling value
Levelling: Temperature gradient still too high•
Levelling demand (after power on): Device in warming phase •
(1 min)
Levelling demand (after power off): Warning for unreliable •
values
2-Sensor-Control:•
Y-axis is more than 5 steps away from target position
Permanently lit
Levelling: X-axis is levelled•
Calibration of X-axis successful•
2-Sensor-Control:•
X-axis completely arranged, target position reached•
5Hz-Flash
Levelling: Y-axis is more than 1000 steps away from calibrated•
levelling value
Levelling: Y-(Z-)axis displays restive values•
Levelling not possible, outside measurement range•
2-Sensor-Control:•
Y-axis is more than 5 steps away from target position
X+
X-
SIGNALS AND SPECIAL FEATURES