10 Alarms and troubleshooting
46
10.3 Time alarm A3 (low priority alarm)
Control panel
Alarm condition
The set temperature (2) is not reached during 60 minutes
of uninterrupted heating.
Device response
A3 (flashing) appears on the display (5).
The Start LED (6) flashes.
The alarm LED (10) lights up yellow.
An acoustic alarm signal is triggered.
The heating process is not interrupted.
Possible reasons
►Required
action(s)
A layer of thermally conductive materials (water or gel) is
situated on the applied part.
►Remove the pad or place it underneath the applied part.
Applied part defective.
►Send the applied part to the local sales point.
Required action(s)
for resetting
Press the Start button (6) to reset the alarm.
If the set temperature is not reached during 60 minutes of
uninterrupted heating, the alarm is triggered again.