EasyManua.ls Logo

STIHLER ELECTRONIC ASTOPAD DUO310 - 12 Brief Overview of Operating States and Alarms; Overview of Operating States

STIHLER ELECTRONIC ASTOPAD DUO310
66 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ASTOPAD
®
Instructions for Use
12 Brief overview of operating states and alarms
52
12 Brief overview of operating states and alarms
12.1 Overview of operating states
Operating
state
Display
Start
LED
Output
A or B
Alarm
LED
Standby
LED
Acoustic
alarm
signal
Possible
reasons
Green
Yellow
Blue
Standby
mode
-
Battery is charg-
ing
Battery defective
or no original
battery inserted
On mode
Battery defec-
tive or no origi-
nal battery in-
serted
No applied part
connected
Actual temperature
Set temperature
Self-test
L
Set temperature
Self-test and
Actual tempera-
ture < 20°C
H
Set temperature
Self-test and ac-
tual temperature
> 45°C
=
LED off
=
LED on
=
LED flashes
11
6
10
9

Table of Contents