14 Technical Data
ASTOPAD DUO310 control unit
Electrical connection
100 240 VAC, 50 60 Hz
Rated current
110 V = 1.6 A, 240 V = 0.8 A
Primary fuses
2 x 3.15 A
Power consumption
max. 160 W
Classification (IEC/EN 60529)
IPX2
Classification (IEC/EN 60601-1)
Protection Class I, defibrillation-protected
applied part type BF
Classification
(MDD 93/42/EEC)
Class IIb
Code UMDNS
10-414(COV), 11-989(SOF/ROE)
Code GMDN
37329
Dimensions (mm)
Height
Width
Depth
max.
300
155
130
Weight (kg)
2.0 (without battery)
2.5 (with battery)
Operating mode
Continuous operation
Permissible environmental
conditions
Moisture
Temperature
Atmospheric
pressure
In operation/storage
10% to 75%
non-condensing
+16°C to +26°C
700 hPa to
1060 hPa
In transport
-20°C to +50°C
Regulation of the contact sur-
face temperature (Essential per-
formance characteristics accord-
ing to IEC/EN 80601-2-35)
32.0°C to 39.0°C in 0.5°C increments
Tolerance ± 1.0°C
Display precision of the contact
surface temperature
± 0.7°C
Overheating shut-off
41.0°C (± 0.5°C)
Acoustic alarm volume level
approx. 60 dB(A)
Expected
operating life
The expected operating life is 10 years from the date of first
use, providing the product is subject to misuse, negligence,
damage or inappropriate use and the equipment is used and
serviced properly and as intended.