ASTOPAD
®
Instructions for Use
57
14 Technical Data
ASTOPAD DUO310 control unit
100 – 240 VAC, 50 – 60 Hz
110 V = 1.6 A, 240 V = 0.8 A
Classification (IEC/EN 60529)
Classification (IEC/EN 60601-1)
Protection Class I, defibrillation-protected
applied part type BF
Classification
(MDD 93/42/EEC)
10-414(COV), 11-989(SOF/ROE)
Dimensions (mm)
Height
Width
Depth
2.0 (without battery)
2.5 (with battery)
Permissible environmental
conditions
10% to 75%
non-condensing
Regulation of the contact sur-
face temperature (Essential per-
formance characteristics accord-
ing to IEC/EN 80601-2-35)
32.0°C to 39.0°C in 0.5°C increments
Tolerance ± 1.0°C
Display precision of the contact
surface temperature
Acoustic alarm volume level
The expected operating life is 10 years from the date of first
use, providing the product is subject to misuse, negligence,
damage or inappropriate use and the equipment is used and
serviced properly and as intended.