19 800 0002 12.2020
19
Operation
CAUTION
If the handpiece is not positioned correctly, there is an
impairment to health due to ineffective treatment!
► Define the treatment zone and make sure that the handpiece position always cor-
responds to the treatment zone.
► Make sure that the treatment is only administered by users who meet the condi-
tions in Chapter 2.2 PRECONDITIONS FOR OPERATION.
For safety reasons, using the device for applications other than those specified in chap-
ter 1 GENERAL SAFETY INFORMATION is not permitted!
CAUTION
Over extended periods, the noise of the pulses
may be perceived as unpleasant!
► Offer ear protection to the patient.
► Recommendation: The user should also wear ear protection.
NOTICE
Danger from inadvertent pulse triggering.
► Only trigger pulses when the handpiece is in contact with the intended treatment
zone.
• Avoid excessive pressure of the pulse transmitter to the patient's skin! Excessive
pressure is not necessary for the success of the treatment.
CAUTION
If too many pulses are applied to the same spot,
there is a risk of swelling, petechiae and haematomas and it could lead to
heating.
► Do not apply more than 300-400 pulses to the same location.
• Apply enough coupling gel to ensure that the pulse transmitter glides smoothly
over the patient's skin.
Setting parameters
Treatment should always start at a low intensity level. This also applies when resuming
treatment after an interruption. The pulse energy should be increased gradually during
treatment.
The selection of intensity levels is based on the medical opinion of the person admin-
istering the treatment. The maximum intensity level used during treatment must not
cause the patient undue pain under any circumstances.