19 800 0002 12.2020
24
Cleaning, Maintenance, Overhaul
Clean the guide tube using a brush.
Follow the instructions in reverse order to assemble the handpiece.
When assembling the handpiece, always retighten the handpiece shaft using the sup-
plied open-end spanner. It must no longer be possible to unscrew the shaft by hand.
5.1.3 Cleaning and disinfecting the pulse transmitter
Unscrew the pulse transmitter screw cap and remove the pulse transmitter insert
from the pulse transmitter screw cap (see Chapter 5.1.1 R
EPLACING THE PULSE
TRANSMITTER
).
Clean all of the parts under running water and dry them.
We recommend that the pulse transmitters be cleaned in an ultrasonic bath.
Disinfect the pulse transmitters and leave them to dry.
Wear gloves.
Only use alcohol-based disinfectants containing aldehyde (not formaldehyde)
for medical devices.
Follow the manufacturer's information regarding the effectiveness and appli-
cation of the disinfectant.