General Safety Information
1.2 Safety during treatment of the patient
General note:
Organs with gas inclusions are NOT allowed to be exposed to pulses.
As it passes through tissue, the pulses' energy is slightly reduced; this reduction is sig-
nificantly weakened by the bone structure.
Pulses can give rise to undesirable reactions. The patient must be continuously ob-
served during the treatment and attention must be paid to any reactions experienced
by the patient. The patient must not be under anaesthetic.
Only perform treatments that have been approved by STORZ MEDICAL AG!
The user is responsible for correctly positioning the handpieces and correctly selecting
the treatment zone.
1.3 Warning against damage to equipment and the device
Any damage to the device resulting from incorrect operation is not covered by the
manufacturer's warranty.
Electromagnetic compatibility
This device complies with the requirements of the applicable standard on electromag-
netic compatibility (EMC). Nevertheless, portable and mobile RF communications
equipment (e.g. mobile phones) can interfere with medical electrical devices.
This device is subject to special precautionary measures regarding EMC and must be
installed in accordance with the EMC directives in Chapter 8.4.1 EMC
GUIDELINES AND
MANUFACTURER
'S DECLARATION.
The use of accessories or cabling not authorised by the manufacturer is not permitted.
Increased electromagnetic interference may occur, which may cause the device to op-
erate incorrectly.
The MASTERPULS
®
ONE is not allowed to be positioned immediately next to or jointly
with other devices. If operation near or jointly with other devices is required, the MAS-
TERPULS
®
ONE must be tested in that particular environment to ensure operation ac-
cording to technical specifications.
The use of RF communication devices in the vicinity is not permitted.
The device must only be connected to properly earthed and correctly installed shock-
proof sockets!
Setup and operation
There are ventilation slits on the rear of the device which must not be covered by other
objects.
• Never cover the device when in use!
• Make absolutely sure that no liquid can seep into the system housing or hand-
piece.
The third mains connecting cable must only be used as the functional earth.