19 800 0002 12.2020
33
Technical Specifications and Conformity
8.4.1 EMC guidelines and manufacturer's declaration
Guidelines and manufacturer's declaration Emitted electromagnetic interference
The MASTERPULS
®
ONE model is intended to be used in the electromagnetic environment specified
below. The customer or the user of the MASTERPULS
®
ONE should ensure that it is used in such an
environment.
The maximum length of the mains cable for the unit is 3 m.
Interference emis-
sion measurements
Compliance
Electromagnetic environment – guidelines
RF emissions acc. to
CISPR 11
Group 1
The MASTERPULS
®
ONE uses RF energy only for its in-
ternal functioning. Therefore, its RF emissions are very
low and are not likely to cause any interference in near-
by electronic equipment.
RF emissions acc. to
CISPR 11
Class B
The MASTERPULS
®
ONE is suitable for use in all estab-
lishments, including domestic establishments and
those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
according to IEC
61000-3-2
Class A
Emissions of voltage
fluctuations/flicker
acc. to IEC 61000-3-
3
Complies
Conducted RF emis-
sions
Class B
Radiated RF emis-
sions
Class B