19 800 0002 12.2020
23
Cleaning, Maintenance, Overhaul
Make sure that the cap parts of the pulse transmitters are screwed firmly in place and
that the pulse transmitter screw cap is screwed firmly to the shaft.
Check the screw connection of the pulse transmitter screw cap and cap parts during
prolonged treatment phases.
5.1.2 Cleaning the handpiece
CAUTION
If the shaft of the handpiece is damaged or uneven, then when removing or screwing
in the shaft without a suitable tool,
minor scratching injuries may occur.
Be sure to use the open-end spanner provided to remove or screw in the shaft.
After cleaning, the handpiece must be dry before it can be reassembled.
Disconnect the handpiece from the control device.
Unscrew the pulse transmitter screw cap from the handpiece.
Clean the coupling gel off the handpiece using a cleaning agent that is suitable for
surface cleaning.
Disinfect the handpiece with an alcohol-based disinfectant containing aldehyde
(follow the manufacturer's instructions).
Unscrew the shaft from the handpiece and pull it out of the handpiece handle.
Use the supplied open-end spanner for this purpose.