LaboPol-35
Instruction Manual
4
Always remember to switch the power off when installing electrical
equipment.
LaboPol-35 is available in 2 voltage variants:
1/3 x 200-240 V / 50-60 Hz
1 x 100 V / 50-60 Hz
The 1/3 x 200-240 V / 50-60 Hz LaboPol-35 is shipped with 2 types
of Mains cables:
The 2-pin (European Schuko) plug is for use on single-phase
connections.
If the plug supplied on this cable is not approved in your country,
then the plug must be replaced with an approved plug. The leads
must be connected as follows:
Yellow/green: earth
Brown: line (live)
Blue: neutral
The 3-pin (North American NEMA 6-15P) plug is for use on 2-phase
power connections.
If the plug supplied on this cable is not approved in your country,
then the plug must be replaced with an approved plug. The leads
must be connected as follows:
Green: earth
Black: line (live)
White: line (live)
Supplying Power
DANGER!
The machine must be earthed
IMPORTANT
Check that the mains voltage corresponds to the voltage stated
on the type plate on the back of the machine.
Connecting a 1/3 x 200-240 V /
50-60 Hz
Single-phase Supply
2-phase Supply