EN-11
Setup and Interconnection
Stryker Endoscopy considers instructional training an integral part of the
Crossre™ system. Your Stryker Endoscopy sales representative will perform at
least one inservice at your convenience to help you set up your equipment and
instruct you and your sta on its operation and maintenance. Please contact
your local Stryker Endoscopy representative to schedule an in-service aer your
equipment has arrived.
Electromagnetic Compatibility
Like other electrical medical equipment, the Crossre™ System requires special
precautions to ensure electromagnetic compatibility with other electrical medical
devices. To ensure electromagnetic compatibility (EMC), the Crossre™ System
must be installed and operated according to the EMC information provided in
this manual.
e Crossre™ System has been designed and tested to comply with IEC 60601-
1-2 requirements for EMC with other devices.
Warning is equipment is intended for use by health care
professionals only. is equipment may cause radio
interference or may disrupt the operation of nearby
equipment. It may be necessary to take mitigation
measures, such as reorienting or relocating the equipment
or shielding the location.
Portable and mobile RF communications equipment
can aect the normal function of the Crossre™ System
even if such equipment meets the applicable emissions
requirements.
e Crossre™ System was not tested for immunity to
electromagnetic disturbances.
Do not use cables or accessories other than those provided
with the Crossre™ System, as this may result in increased
electromagnetic emissions or decreased immunity to such
emissions.