Technical Specifications
Stryker Endoscopy reserves the right to make improvements to the product(s)
described herein. Product(s), therefore, may not agree in detail to the published
design or specications. All specications are subject to change without notice.
Please contact the local Stryker Endoscopy distributor or call your local Stryker
Endoscopy sales representative or agent for information on changes and new
products.
Dimensions
Size: 16.9" L × 12.5" H × 4.5" W
Weight: 20 lbs
Environmental Specifications
Operating temperature: 5 – 40°C
Operating humidity: 30 – 95% RH
Shipping temperature: -18 – 60°C
Shipping humidity: 15 – 90% RH
System Input Power Requirements
Voltage: 100-240 VAC @ 50/60Hz, 6 – 10 A
Inlet Fuse: F 16.0AH, 250V
Electrical Specifications
Motor output max speed: 12000 RPM
Motor duty cycle: Continuous operation
RF output waveform: 200 kHz ± 1%, square wave,
Crest factor <1.5 @ 200 ohms
Generator Output
Output power at each set point with specied load resistance (per IEC 60601-2-2,
sub clause 6.8.3) is given in the graphs below.