EN-27
Using the Handpiece
Warning During use, the RF and shaver handpieces generate
electronic noise that may interfere with EKG readings.
Before responding to any erratic EKG readings, rst power
down the system to ensure the readings are not the result of
system noise.
RF handpieces are intended for single use only and should
not be reprocessed or reused.
Shaver handpieces are provided nonsterile and must
be cleaned and sterilized prior to each use, according to
the reprocessing instructions provided in the handpiece
manual.
Each handpiece has its own set of controls.
RF Hand Controls
1
2
3
1. Adjust CUT power level (single press)
or
Activate/deactivate Force Modulation
(press and hold for three seconds)
2. Activate CUT
3. ActivateCOAG