EN-4
3. Electrosurgical components, such as the probe, may remain hot
following activation. Keep all electrosurgical equipment away from
ammable materials to avoid combustion.
4. To prevent the risk of re, DO NOT replace console fuses. If it is
suspected that fuses are damaged, return console to Stryker for repair.
5. Flammable agents used for cleaning or disinfecting, or as solvents of
adhesives, should be allowed to evaporate before using the Crossre™
system. ere is a risk of pooling of ammable solutions under the
patient or in body depressions such as the umbilicus, and in body
cavities such as the vagina. Any uid pooled in these areas should be
mopped up before Crossre™ system is used. Attention should be called
to the danger of ignition of endogenous gases. Some materials, for
example cotton and gauze, when saturated with oxygen may be ignited
by sparks produced in normal use of the Crossre™ system.
Prior to Surgery
1. e operator of the Crossre™ system should be a qualied physician,
having complete knowledge of the use of this equipment and awareness
of the risks associated with arthroscopic and electrosurgical procedures.
2. e operator of the Crossre™ system should be experienced in
arthroscopic and electrosurgical practices and techniques.
3. e operator of the Crossre™ system should read this manual
thoroughly and be familiar with its contents prior to operating the
equipment.
4. e operator of the Crossre™ system should be sure that the system
functions as outlined in this manual prior to a surgical procedure. e
Crossre™ system was fully tested at the factory before shipment.
5. Crossre™ system components are designed to be used together as a
system. Use only the appropriate footswitch, handpiece, and disposable
attachments described in this manual.
6. Carefully unpack the unit and ensure that all components are accounted
for and remain undamaged from shipment. Inspect the handpiece cable
for any damage to insulation. If damage to any component is detected,
refer to the “Service and Claims” section of this manual.
7. Ensure the proper connection of the primary power cord of the
Crossre™ System to a grounded receptacle. To prevent risk of electric
shock DO NOT use extension cords or adapter plugs.
8. DO NOT wrap the handpiece cable around metal objects, or the
induction of hazardous currents may result.
9. Position the cables to avoid contact with the patient, electrodes, cables,
and any other electrical leads which provide paths for high frequency