EN-6
10. Ensure that the probe tip, including the return electrode, is completely
surrounded by irrigant solution during use.
11. Keep the activation indication lights and speaker in eld of view and
hearing at all times during activation. e light and sound are important
safety features.
12. DO NOT touch the attachment to metal objects, such as an endoscope
or metal cannula, while activating the handpiece. Damage to the
attachments or other devices may result.
13. DO NOT obstruct the fan (located near the rear of the console).
14. Failure of the system may result in an unintended increase in output
power.
15. During use, operators should wear standard surgical gloves to help
reduce the risk of electric shock.
After Surgery
1. DO NOT attempt to reuse or resterilize any product labeled “Single-
Use,” as this may lead to equipment malfunction, patient/user injury,
and/or cross contamination.
2. DO NOT use ammable agents for cleaning and disinfection of the
Crossre™ console, handpiece, or footswitch.
3. DO NOT remove the cover of the console as this could cause electric
shock and product damage.
4. Attempt no internal repairs or adjustments, unless specied otherwise
in this manual. Units requiring repair should be returned to Stryker.
5. Disconnect the Crossre™ system from the electrical output when
inspecting fuses.