Scale System Operation
5
1.
3.
2.
4.
Battery Charge Level Indicator
1. Display − Displays patient weight, unit of measurement and battery status.
2. “Zero” − Push and hold for 2 seconds to zero the scale system before putting a patient on the stretcher.
If the display flashes “hold”, press and hold the “Zero” button again until the display reads “rEL” (release).
Release the “Zero” button. The display flashes “000.0”, then displays “000.0”. The system is not zeroed
until the “000.0” stops flashing. For the most accurate results, always zero the scale system before putting
a new patient on the stretcher. The display will shut off after approximately 40 seconds.
3. “Weigh” − Push to weigh the patient. The display will show the patient’s weight for approximately 40
seconds before turning off.
4. “lb/kg” − Push to display patient weight in pounds or kilograms.
NOTE
Do not touch the stretcher while the scale system is weighing or zeroing.
The patient must remain still while the system is weighing. If the patient is moving, the system will try for 20
seconds to get a stable weight or zero value before displaying the error message .
If there is a loose connection or a malfunctioning component, the display will show “Err”. Attempt the function
again. If the system is functional, “Good” will display and the scale system is ready to use. If the malfunction
is still present, the display shows “Err” again. Call Stryker technical support at 800−327−0770.
For the most accurate results, weigh the patient with the litter at zero degrees of Trend.
SYMBOL ACTION DISPLAY
Press and release ”WEIGH”. − − − −
"XXX.X lbs
Press and hold ”ZERO” "hold
"rEL"
Release ”ZERO” "000.0" (flashing)
"000.0 (solid)
To convert the patient’s weight "XXX.X kg
to kilograms, press and release
”lb/kg”
Repeat to return to pounds. XXX.X lbs
Weigh
Zero
lb/kg