Operative Technique
Using the Recon Screwdriver the
correct Lag Screw is inserted through
the Tissue Protection Sleeve and
threaded up to the subcondral part
of the femoral head. The screw is
near its proper seating position when
the groove around the shaft of the
screwdriver is approaching the end of
the Tissue Protection Sleeve (Fig. 32,
32a). The required length of the
second Lag Screw can be measured
using the Recon Lag Screw Gauge.
Remove the Distal K-Wire and K-Wire
Sleeve. Then insert the Sleeve for the
solid Stepdrill into the distal Tissue
Protection Sleeve.
Repeat the same surgical steps for
drilling and insertion of the distal Lag
Screw without K-Wire guidance.
After the completion of the distal Lag
Screw insertion, move on to the distal
locking procedure.
Fig. 32
Fig. 32a
25