Calibration-213
Service
Description
1. Digital
Voltmeter
1
2.
DC
Voltage
Standard
1
3.
DC
Current
Standard
4.
Resistance
Standard
1
5. AC
Calibration
Stan-
dard
3
6. Sine-Wave
Generator
1
7.
Amplitude
Calibrator
1
8. Time-Mark
Generator
1
9. Screwdriver, Low Capa-
citance
10. Screwdriver
TABLE
6-1
Equipment Required
Minimum
Specifications Purpose
Range, 0.1
to
1000
volts
de
and Power supply
check
and
true
rms; accuracy,
within
0.1%
adjustment;
rms conver-
from
de
to
1 kHz and 0.6%
at
ter
check
and
adjustment;
40
kHz.
battery
charger check;
and
crt
grid bias
check
and ad-
justment.
Range, 0.1
to
800
volts; accu-
De voltage accuracy check
racy,
within
0.01%. and
adjustment;
overrange
check;
character
generator
check; rms
converter
de
response
check.
Range, 0.1
to
1000
mA;
accu-
De
current
measurement
racy,
within
0.02%. accuracy check.
Range,
500
n
to
5
MD.
accu-
Resistance check and ad-
racy,
within
0.05%.
justment.
Range,
10
mV
fo
500
V; fre-
Ac voltage and
current
quency
output,
1 kHz
to
at
Measurement accuracy
least 1 MHz; accuracy,
within
check.
0.5%
for
voltage and
2%
for
frequency.
Frequency
Range, 1 kHz
to
at
Bandwidth check; trigger
least 1 MHz;
output,
0.1
to
at
level and slope check and
5.5
V p-p.
adjustment;
rms
converter
check
and
adjustment.
Output,
20
mV
to
100
Vat
1 Vertical gain check and ad-
kHz; accuracy, 0.3%.
justment;
attenuator
com-
pensation
check
and
adjust-
ment;
variable vertical gain
check.
Marker
outputs,
2 µs
to
0.2
s
Sweep
timing
check
and ad-
in
1-2-5 sequence; accuracy,
justment;
horizontal mag-
within
0.1%.
nifier
check;
sweep lineari-
ty
check.
Bit size,
3/32
inch;
shank,
at Used
throughout
procedure
least 3 inches.
to
adjust
variable capacitors.
Bit size,
3/32
inch;
shank,
at
Used
throughout
procedure
least 3 inches.
to
adjust
variable
compo-
nents.
1
Required
for
calibration adjustments.
Other
equipment
is
required
to
check performance specifications.
2
Requires a
TM
500
series
Power
Module.
Example
of
Applicable
Test
Equipment
Fluke Model
8375A
Digital
Voltmeter.
Fluke
Model
341A
or
343A
Voltage Calibrator.
Fluke Model
382A
Voltage/
Current
Calibrator.
Electro Scientific Industries
Model DB62
Dekabox.
Optimation
Model AC-126
AC-DC Voltage Calibrator
with Model PA-226 Power
Amplifier and Model CS-11-
/2206
Current
Calibrator.
General Radio Model
1310A
or
1310B Oscillator.
Tektronix
Part
No.
067-
0502-01,
Standard
Ampli-
tude
Calibrator.
a.
Tektronix
TG 501 Time-
Mark
Generator
2
•
b.
Tektronix
2901 Time-
Mark
Generator.
JFD
Electronics
Corp
ad-
justment
tool,
Number
5284
(Tektronix
Part No. 003-
0489-00).
Xcelite R-3323
(Tektronix
Part
No.
003-0192-00).
3
This
is
a high cost item which
is
required
to
verify
the
accuracy
of
the
213
for
ac
(rms) voltage and current measurements.
It
is
not
required
to
perform
the calibration adjustments. Refer
to
the
Check
RMS
Voltage Measurement Accuracy calibration procedure
for
further
information.
6-2
@