Cal 
ibration-213 
Service 
d.  Set 
the 
Standard 
output 
to 
+1.000 
V. 
e. 
CHECK-Reading 
of +1.000 ± 
0.025 
(0.975 
to 
1.025). 
f. 
Disconnect 
the 
Standard. 
12.  CHECK RMS 
VOLTAGE 
MEASUREMENT 
ACCURACY 
NOTE 
The 
AC 
Calibration Standard 
is 
required 
to 
verify 
the specified accuracy 
of 
each  rms 
voltage range 
with 
this procedure (step  12). 
See 
footnote 
3 
in 
Table 6-1. 
If 
an 
AC 
Cali-
bration 
Standard is 
not 
available, the 
follow-
ing procedures are 
provided 
as 
alternatives to 
the 
procedure 
following 
this note. 
(a) 
If 
all 
previous calibration steps have been 
accomplished, each 
circuit 
function 
involved 
in rms voltage measurement has been checked 
and 
adjusted; therefore, 
further 
rms voltage 
measurement accuracy checks 
may 
not 
be 
necessary 
for 
the instruments 
intended 
use. 
(b)  The procedure 
and 
equipment 
used 
in 
Calibration 
step  10 can also be used 
to 
verify 
the 
0. 
1, 
1, 
10, 
and 
100 V rms 
ranges. 
The 
1000 V range can 
be 
checked, 
but 
the 
output 
of 
the Sinewave Generator 
is 
not 
sufficient 
to 
provide 
the 
minimum 
10% 
of 
full 
scale 
input 
to 
verify 
the performance specification. 
(c)  Calibration step  12A 
is 
an alternate 
method 
of 
verifying the 
0. 
1, 
1, 
10, 
and 
100 V 
rms 
ranges_ 
The  1000 V range also can be 
checked 
with 
this alternate procedure, 
but 
the rms value 
of 
the  100 V 
fixed 
output 
of 
the 
Amplitude 
Calibrator 
is 
not 
sufficient 
(at least 
10% 
of 
full 
scale)  to 
verify 
the per-
formance specification. 
a.  Set INPUT COUPLING 
to 
AC. 
b.  Connect 
the 
probe 
to 
the 
AC 
Calibration Standard 
(Voltage configuration) with a dual  banana-plug 
to 
BNC 
adapter and a  probe adapter. 
c. 
Set VOL 
TS-mA/DIV 
and 
the 
Standard 
as 
show 
in 
Table 6-9. 
6-16 
d. 
CHECK-Reading 
is 
within limits shown in Table 6-9. 
e. 
Disconnect 
the 
Standard. 
TABLE 6-9 
RMS Voltage Measurement Accuracy 
VOLTS-
AC 
Calibration 
mA/DIV 
Standard 
Limits 
Setting 
Setting 
(blue 
(set 
at 
4 
kHz 
4 
kHz 
40 
kHz 
scale) 
and 
40 
kHz) 
Reading 
0.1 
0.0100 
.0100 
.0093 
to 
.0107 
.0091 
to 
.0109 
0.2000 
.2000 
.1965 
to 
.2035 
.1925 
to 
.2075 
1 
0.100 
0.100 
0.094 
to 
0.106 
0.094 
to 
0.106 
2.000  2.000 
1.975 
to 
2.025 
1.975 
to 
2.025 
10 
1.000 
01.00 
00.94 
to 
01.06 
00.94 
to 
01.06 
20.00 
20.00 
19. 
75 
to 
20.25 
19. 75 
to 
20.25 
100 
10.00 
010.0 
009.4 
to 
010.6 
009.4 
to 
010.6 
200.0 
200.0 
197.5 
to 
202.5 
197 .5 
to 
202.5 
1000 
100.0 
0100 
0094 
to 
0106 
0093 
to 
0107 
500.0 
0500 
0490 
to 
0510 
0485 
to 
0515 
12A. 
ALTERNATE 
CHECK OF RMS 
VOLTAGE 
MEASURE-
MENT 
ACCURACY 
NOTE 
This procedure assures the specified accuracy 
of 
the 213, 
but 
does 
not 
check 
all 
ranges 
(see 
Note 
in 
Calibration step  12). 
a. 
Connect 
the 
Digital 
Voltmeter 
and the 
probe 
to 
the 
Amplitude Calibrator with a T-Connector, dual  banana-
plug 
to 
BNC 
adapter, probe adapter, and a cable.  Set 
the 
Calibrator for a square wave 
output. 
b.  Set  INPUT COUPLING 
to 
AC. 
c. 
Set VOL TS-mA/DIV and  Calibrator 
as 
shown 
in 
Table 
6-10. 
d. 
CHECK-Reading 
is 
within limits shown 
in 
Table 6-10. 
e. 
Disconnect 
the 
Voltmeter and Calibrator. 
@