Getting Acquain
tedwithYourInstrument
Quick Tips
Each channel has independent output on and off control. When the output status of Ch n is set to On, both analog output
and marker output are enabled.
When Direct Output is set to On, filter and offset settings become inactive. The analog bandwidth setting range will
change.
(AWG7000 series) The Option 02 (Extended analog output bandwidth) and O ption 06 (Interleave and wide bandwidth
direct output) models do not have the control settings for Filter, Offset, and Direct Output.
(AWG5000 series) You can use Add Input when Direct Output is set to off. If Direct Output is selected, Add Input
is disabled.
The Waveform Rotation can be set independently for each channel.
Analog Phase – Can be set by degree (°).
Analog Delay – Can be set by time or points.
The Waveform Rotation setting does not affect the Waveform window display.
DC Output Page
1. The arbitrary waveform generator has
four lines of DC output.
The level of each output can be set
independently.
2. The output state (On or Off) of the
DC output is common to all DC output
channels.
AWG5000 and AWG7000 Series Quick Start User Manual 33