Acquiring Waveforms
CSA7000 Series, TDS7000 Series, & TDS6000 S eries Instruments User Manual
3-57
FastFrame is not compatible with these featur es or modes:
H Equivalent Time
H Histograms
H Fast Acquisitions
H Average
H Envelope
H Waveform Database
Consider the following operating characteristics when using FastFrame:
H You can push RUN/STOP to terminate a FastFrame sequence. If any frames
were acquired, they are displayed. If no frames were acquired, the previous
FastFrame waveform is displayed.
H Because FastFrame introduces additional processing time into the operation
cycle of acquire, process, and display, its best to use Single Sequence
Acquisition (see Acquire menu, S top After menu). With Single Sequence
selected, you will see the current acquisition sequence; otherwise, the display
lags the current sequence by one sequence. You can also see the current
sequence by pushing the RUN/STOP button to stop the acquisition.
Using FastFrame
Acquisitions