Acquiring Waveforms
3-62
CSA7000 Series, TDS7000 Series, & TDS6000 S eries Instruments User Manual
Overview Control elements and resourcesTime stamping frames (Cont.)
To lock the
reference
position
frames
10. Touch the Horiz button. Select the Acquisition tab from
the Horiz/Acq control window. Touch FastFrame Setup
to display the FastFrame Setup control window.
Note. You can also get the FastFrame Setup control
window by pushing the Set Up button on the Selection
Controls window.
11. Touch either Frame Tracking Live or All to lock the
reference and position frames together. When the
frames are locked, they maintain the same relative
distance from each other when either frame is adjusted:
H Live locks together the channel and math
waveforms. All reference waveforms are locked
together, but they are separate from the channel
and math waveforms.
H All locks together all channel, math, and reference
waveforms; adjusting one waveform adjusts all
waveforms.
Time difference between the reference
and selected frames
Trigger time of the reference frame
Trigger time of the selected frame
Figure 3- 17: FastFrame time stamp