Page #175 background image
Triggering
CSA7000 Series, TDS7000 Series, & TDS6000 S eries Instruments User Manual
3-97
Overview Control elements and resourcesTo trigger on a runt pulse (Cont.)
Set to trigger
when
To determine how wide a runt pulse the instrument will trigger
on:
5. Touch Trigger When Runt and select from the list:
H Occurs triggers on all runt pulses regardless of
width.
H Wider triggers only on runt pulses that exceed the
width you set. Enter the width using the general
purpose knob or keypad.
6. To specify the minimum width of the runt pulse, touch
Width, and set the value using the multipurpose knob or
keypad.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix CSA7404 and is the answer not in the manual?

Tektronix CSA7404 Specifications

General IconGeneral
Bandwidth4 GHz
Sample Rate20 GS/s
Input Channels4
Vertical Resolution8 bits
Input CouplingAC, DC, GND
Timebase Accuracy± 50 ppm
Power Supply100-240 VAC, 50/60 Hz
Display10.4 inch color LCD
Record Length1 M points
Trigger ModesEdge, Glitch, Width, Timeout
InterfaceUSB, GPIB, Ethernet
Weight40 lb (18.2 kg)