EasyManuals Logo
Home>Tektronix>Measuring Instruments>CSA7404

Tektronix CSA7404 User Manual

Tektronix CSA7404
480 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #175 background imageLoading...
Page #175 background image
Triggering
CSA7000 Series, TDS7000 Series, & TDS6000 S eries Instruments User Manual
3-97
Overview Control elements and resourcesTo trigger on a runt pulse (Cont.)
Set to trigger
when
To determine how wide a runt pulse the instrument will trigger
on:
5. Touch Trigger When Runt and select from the list:
H Occurs triggers on all runt pulses regardless of
width.
H Wider triggers only on runt pulses that exceed the
width you set. Enter the width using the general
purpose knob or keypad.
6. To specify the minimum width of the runt pulse, touch
Width, and set the value using the multipurpose knob or
keypad.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix CSA7404 and is the answer not in the manual?

Tektronix CSA7404 Specifications

General IconGeneral
BrandTektronix
ModelCSA7404
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals