Commands Listed in Alphabetical Order
DISplay:MEAS
Readout
This command sets or queries state of the measurement readout. This is equivalent
to toggling the Measurement readout off or on with the M easurement readout
icon in the to
olbar.
Group
Display Control
Syntax
DISplay:MEASReadout { ON | O FF | NR1 }
DISplay:MEASReadout?
Arguments
ON enables the readout display for the Measurement function.
OFF disables the readout display for the Measurement function.
NR1 set to 0 disables the Measurement readout d isplay; any other integer
enables the Measurement readout display.
Examples
DISPLAY:MEASREADOUT 1 enables the Measurement readout display.
DISPLAY:
MEASREADOUT?
might return DISPLAY:MEA SREADOUT 0, indicating
that the Measurement readout display is off.
NOTE. A query always returns a 0 or 1.
DISplay:PERSistence
This c
ommandsetsorqueriesthelengthoftimethatdatapointsaredisplayed
when variable persistence is a ctive. This affects the display only and is equivalent
to setting Time in the Style section of the Display Setup dialog box.
Group
Display Control
Syntax
DISplay:PERSistence <NR3>
DISplay:PERSistence?
Related Commands
DISplay:STYle
Arguments
NR3 specifies the length, in seconds, that the waveform points are displayed on the
screen. The range is 500 ms to 100 s (in increments of 100 ms from 500 ms to 1s
and i
n 0.5 s increments from 1 s to 100 s).
DSA/CSA/TDS8X00/B Series Programmer Manual 2-129