EasyManuals Logo

Tektronix DSA8200 Series User Manual

Tektronix DSA8200 Series
431 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #214 background imageLoading...
Page #214 background image
Commands Listed in Alphabetical Order
Arguments
NR1 is an intege
r in the range of 0 through 15 representing a color as defined
in the above table.
Examples
MASK:COLOR 4 sets the color of the mask to fuchsia.
MASK:COLOR? might return MASK:COLOR 10, indicating that the color of the
mask is green.
MASK:COUNt
This command clears the counts for the masks, clears the data source, and restarts
the counting. This is equivalent to clicking Clear in the Mask Counts section
of the Mask Setup dialog box. The query form of this command returns all the
values for the mask count parameters.
Group
Mask
Syntax
MASK:COUNt {RESET}
MASK:COUNt?
Examples
MASK:COUNT RESET clears the counts for the masks, clears the data source,
and restarts the counting.
MASK:COUNT? might return MASK:COUNT:STATE 1;TOTAL
487641;WAVEFORMS 2887;SAMPLES 1443500
.
MASK:COUNt:SAMPles? (Query Only)
This query only command returns the total number of sample points used in the
count of the total number of mask hits (returned by the MASK:COUNt:TOTal?
query).
Group
Mask
Syntax
MASK:COUNt:SAMPles?
Examples
MASK:COUNT:SAMPLES? might return MASK:COUNT:SAMPLES 50 0, indicating
that the count of mask hits is based on 500 samples.
2-200 DSA/CSA/TDS8X00/B Series Programmer Manual

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA8200 Series and is the answer not in the manual?