Commands Listed in Alphabetical Order
Examples
MEASUREMENT:M
EAS2:SOURCE1:EDGE:DIRECTION FORWARD
causes the
instrument to search forward from the beginning of the Source 1 waveform (or
the beginning of the gated region, if gates are enabled) for the rst specied edge
(rising or falling).
MEASUREMENT:MEAS8:SOURCE1 :EDGE:DIRECTION? might return
MEASUREMENT:MEAS8:SOURCE1 :EDGE:DIRECTION FORWARD, indicating that
the instrument will search forward from the beginning of the Source 1 waveform
(or the beginning of the gated region, if gates are enabled) for the specied edge
(rising or
falling) when taking Measurement 8 .
MEASUrement:MEAS<x>:SOUrce<x>:EDGE:SLOPe
This command sets or queries the slope (or polarity) of the edges used in delay
time measurements. The measurement slot is specied by x, which can be 1
through 8; the source is specied by x, which can be 1 (source waveform) or
2 (destination waveform). For edge measurements other than delay, the slope
is dete
rmined implicitly (for example, rise time measurements use positive
slopes). If gates are enabled (see the MEASUrement:MEAS<x>:GATing:STATE
command), the instrument nds the edge within the measurement zone; o therwise,
the instrument nds the edge within the entire waveform record.
Sending this command is equivalent to selecting an option button (±, +, or -) for
Slope in the Edges section of the Region dialog in the Measurement Setup dialog
box. (A delay measurement must be selected before these buttons are active.)
Use the MEASUrement:MEAS<x>:SOURCE:EDGE:DIRection command to
set the direction that the instrument uses to look for the rising or falling edge
of the waveforms.
NO
TE. The SOURCE<x> SOURCE:<<x> argument species the
source, which can be source 1 or source 2. To set the actual waveform
(channel, reference, or math) that is assigned to source, you must use the
MEASUrement:MEAS<x>:SOURCE<x> WFM command. Two sources are
provided because some measurements (gain, phase) require two sources.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SOUrc e<x>:EDGE:SLOPe {RISE | FALL |
EITHer}
MEASUrement:MEAS<x>:SOUrc e<x>:EDGE:SLOPe?
2-238 DSA/CSA/TDS8X00/B Series Programmer Manual

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA8200 Series and is the answer not in the manual?

Tektronix DSA8200 Series Specifications

General IconGeneral
BrandTektronix
ModelDSA8200 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals