Commands Listed in Alphabetical Order
REF<x>:WFMLabel
This command sets or queries the label associated with the reference waveform
specified.
Group
Vertical
Syntax
REF<x>:WFMLabel <Qstring>
REF<x>:WFMLabel?
Related Commands
CH<x>:WFMLabel, MATH<x>:WFMLabel
Arguments
Qstring sets the label for the reference waveform.
Examples
REF1:WFMLABEL "MY REF1 WAVEFORM" sets the label for REF1 to be "My
REF1 waveform," which is the label d isplayed with the waveform when it is
displayed on s creen.
REF1:WFMLABEL? might return REF1:WFMLABEL "MY REF1 WAVEFORM",
indicating that the label for the REF1 waveform is set to "My REF1 waveform."
*RST (No Query Form)
This command (no query form) resets the instrument to the factory default settings.
The *RST command does not alter the following:
The state of the IEEE Std 488.1-1987 interface
The selected IEEE Std 488.1-1987 address of the instrument
Calibration data that affect device specifications
The Output Queue
The Service Request Enable R egister setting
The Standard Event Status Enable Register setting
The Power-on status clear flag setting
Stored settings
This command is equivalent to pressing the DEFAULT SETUP buttononthe
front panel.
Group
Status and Error
2-272 DSA/CSA/TDS8X00/B Series Programmer Manual