Commands Listed in Alphabetical Order
Syntax
*RST
Related Commands
FACtory, RECAll:SETUp, SAVe:SETUp
Examples
*RST resets the instrument s ettings to factory defaults.
SAVe:SETUp (No Query Form)
This command (no query form) saves the current instrument setup into the
specified memory location. Sending this is equivalent to selecting Save Setup
in the File menu.
Group
Save and Recall
Syntax
SAVe:SETUp <file path>
Related Commands
FACtory, RECAll:SETUp
Arguments
<FILE P ATH> is a quoted string that defines the file name and p ath. Input the file
path using the form <drive>/<dir>/<filename>. <drive> and one or more <dir>s
are optional. If you do not specify them, the instrument will write the file to the
current directory. The <filename> can be a Windows long file name. Do not use
wild card characters. It is recommended that you use "STP" for the extension
to identify files that store setup data.
Settings saved in one instrument may or may not work on a different instrument
with a different version of firmware.
Examples
SAVE:SETUP "C:\MY DOCU MENTS\TESTS\UI\D ATA\SETUPTEST.S TP" saves
the cu
rrent instrument setup to the file Setuptest.STP in the Data subdirectory on
the C drive.
SAVe:WAVEform (No Query Form)
This command (no query form) stores a waveform in one of eight reference
memory locations or to a file. You must specify a waveform, a timebase, and
a destination (either a file path or a waveform reference slot). Sending this
co
mmand is equivalent to selecting Save Waveform... from the File menu and
then selecting a waveform and destination in the Save Waveform(s) dialog box.