Commands Listed in Alphabetical Order
FACtory (No Qu
ery Form)
This command (no query form) resets the instrument to its factory default settings.
Sending this command is equivalent to pressing the front-panel DEFAULT
SETUP button
or selecting Default Setup in the Utilities menu.
Executing the
FACtory command does the following actions:
Clears the Event Status Enable Register
Clears the Service Request Enable Register
Sets the Device Event Status Enable Register to 25
5
Sets the Power On Status Clear Flag to TRUE
Enables all Command Headers (HEADer ON)
Clears the pending operation flag and associated operations
Clears all references from the reference slots
The FACtory command does not change the following:
The state of the GPIB (IEEE Std 488.1-1987) interface
The selected GPIB address
Data transfer settings
Calibration data that affects device specifications
Protected user data
Stored settings
Group
Miscellaneous
Syntax
FACtory
Related Commands
RECAll:SETUp, *RST, SAVe:SETUp
Examples
FACTORY resets the instrument to its factory default settings.
FILESystem:READFile? (Query Only)
Outputs the specified file to the GPIB port with maximum file size of 10 Mb
Group
Miscellaneous
2-136 DSA/CSA/TDS8X00/B Series Programmer Manual