Commands Listed in Alphabetical Order
Examples
SYSTEM:PROPER
TIES:CH3:IMPEDANCE?
might return
SYSTEM:PROPERTIES:CH3:IMPE DANCE 5.00000000000E+001,
indicating that the characteristic input impedance is 50 for Channel 3.
SYSTem:PROPerties:CH<x>:MODElnum? (Query Only)
This query only command returns the sampling module model number of the
specied channel.
Group
System
Syntax
SYSTem:PROPerties:CH<x>:MO DElnum?
Examples
SYSTEM:PROPERTIES:CH4:MODE LNUM? might return
SYSTEM:PROPERTIES:CH4:MODE LNUM "80E04".
SYSTem:PROPerties:CH<x>:NONDestruct? (Query Only)
This query only command returns the lower (minimum) and upper (maximum)
absolute input signal nondestructive limits of the channel specied by x, which
can be 1 through 8.
Group
System
Syntax
SYSTem:PROPerties:CH<x>:NO NDestruct?
Examples
SYSTEM:PROPERTIES:CH3:NOND ESTRUCT? might
return
SYSTEM:PROPERTIES:CH3 NONDESTRUCT
–3.00000000000E+000,3.0000 0000000E+000
, indicating that the absolute
input nondestruct range for Channel 3 is from –3.0 to +3.0 (with the a pplicable
unit of measure determined by the type of sampling module for the channel).
SYSTem:PROPerties:CH<x>:OPERating? (Query Only)
This query only command returns the lower (minimum) and upper (maximum)
absolute input signal operating limits of the sampling module channel specied by
x, which can be 1 through 8.
Group
System
DSA/CSA/TDS8X00/B Series Programmer Manual 2-285

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA8200 Series and is the answer not in the manual?

Tektronix DSA8200 Series Specifications

General IconGeneral
BrandTektronix
ModelDSA8200 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals