Commands Listed in Alphabetical Order
Syntax
SYSTem:PROPer
ties:CH<x>:OPERating?
Examples
SYSTEM:PROPERTIES:CH5:OPE RATING? might return
SYSTEM:PROPERTIES:CH5:OPE RATING –1.600000000000E+0 00,
1.600000000000E+000
, indicating that the absolute input operating range for
CH5isfrom–1.6to+1.6(withtheapplicableunitofmeasuredeterminedbythe
type of sampling module for the channel).
SYSTem:PROPerties:CH<x>:PRAnge? (Query Only)
Like the PHAseref:CH<x>:RANge? query, this query returns the guaranteed
frequency range of the Phase Reference Module corresponding to CH<x>,
but it returns the range as part of the system-properties query results. See the
PHAseref:CH<x>:RANge? query for more information.
Group
System
Syntax
SYSTem:PROPerties:CH<x>:P RAnge?
Related Commands
PHAseref:CH<x>:RANge?, PHAseref:CH<x>:STAtus?
Examples
SYSTEM:PROPERTIES:CH2:PRA NGE? might return 8.000GHZ-60 .00GHZ for
an 82A04 module installed in the C1C2 small-module compartment.
SYSTem:PROPerties:CH<x>:PRObe:DYNamic? (Query Only)
This query only command returns the lower (minimum) and upper (maximum)
output signal dynamic limits of a probe attached to the channel spe cified by x,
which can be 1 through 8. The actual input dynamic range of the probe (and
channel) can be calculated by dividing the dynamic range limits of the probe by
the scale factor of the probe.
Group
System
Syntax
SYSTem:PROPerties:CH<x>:P RObe:DYNamic?
Related Commands
SYSTem:PROPerties:CH<x>:PRObe:SCAle?
2-286 DSA/CSA/TDS8X00/B Series Programmer Manual