Commands Listed in Alphabetical Order
SYSTem:PROPe
rties:CH<x>:PSYNc:PLENRAnge? (Query Only)
This query returns the guaranteed pattern length range of the PatternSync Trigger
module corresponding to CH<x>, where <x> is a channel associated with the
compartment
in which the PatternSync Trigger module is installed. Use the odd
numbered channel associated with the compartment. For example, if the module
is installed in the CH7/CH8 compartment, use 7 for x.
Group
System
Syntax
SYSTem:PROPerties:CH<x>:PS YNc:PLENRAnge?
Examples
SYSTEM:PROPERTIES:CH1:PSYN C:PLENRANGE? might return
SYSTEM:PROPERTIES:CH1:PSYN C:PLENRANGE "1BIT-8388608BITS" for
an 80A06 PatternSync module installed in the C1C2 small-module compartment.
SYSTem:PROPerties:CH<x>:RISetime? (Query Only)
This query only command returns the characteristic risetime of the sampling
module (if applicable) for the channel specied by x, which can be 1 through 8.
Group
System
Syntax
SYSTem:PROPerties:CH<x>:RI Setime?
Examples
SYSTEM:PROPERTIES:CH4:RISE TIME? might return
SYSTEM:PROPERTIES:CH4:RISE TIME 1.74999997132E-011, indicating that
the characteristic risetime for Channel 4 is 17.5 ps.
SYSTem:PROPerties:CH<x>:SERialnum? (Query Only)
This query only command returns the module serial number of the channel
specied by x, which can be 1 through 8.
Group
System
Syntax
SYSTem:PROPerties:CH<x>:SE Rialnum?
DSA/CSA/TDS8X00/B Series Programmer Manual 2-289