Commands Listed in Alphabetical Order
"1BIT-8388608
BITS"
for a PatternSync Trigger module connected to the
TRIGGER PROBE POWER connector.
TDR? (Query Only)
This query only command returns all of the TDR parameters.
Group
TDR
Syntax
TDR?
Examples
TDR? might return T DR:CH1:UNITS RHO; STEP:STATE 0; DESKEW
0.00000000000; POLARITY PLUS; :TDR:CH2:UNITS RHO; STEP:STATE
0; DESK EW 0.00000000000; POLARITY PLUS; :TDR:CH3:UNITS
RHO; ST
EP:STATE 0;DESKEW 10.500 00000000;POLARIT Y
MINUS; :TDR:CH4:UNITS OHM; STEP:STATE 1;DESKEW
-11.00000000000;POLARITY MINUS; :TDR:CH5:UNITS RHO;
STEP:STATE 0; DESKEW 0.00000000000; POLARITY PLUS;
:TDR:CH6:UNITS RHO; STEP:STATE 0; DESKEW 0. 00000000000;
POLARITY PLUS; :TDR:CH 7:UNITS RHO; STEP:STATE 0;DESKEW
0.00
000000000; POLARITY PLUS; :TDR:CH8:UNITS RHO; STEP:S TATE
0; DESK EW 0.00000000000; POLARITY PLUS; :TDR:INTRATE 100
.
TDR:CH<x>:DIFfpreset (No Query Form)
This command (no query form) sets the instrument to predefined differential TDR
settings for the channel pair specified by x, which can be 1 through 8. A channel
pair are the channels assigned to each small compartme nt. You ca n specify either
c
hannel of a channel pair.
DSA/CSA/TDS8X00/B Series Programmer Manual 2-295