Commands Listed in Alphabetical Order
Syntax
TRIGger:AUTOS
ync {EXEC ute}
Arguments
EXECute executes the AutoSync operations.
Examples
TRIGGER:AUTOSYNC EXECUTE executes all the AutoSync operations.
TRIGger:AUTOSync:DATARate
This command sets the AutoSync data rate selection option to ON or OFF, so
that when AutoSync executes it will or will not synchronize the data rate to the
currently selected waveform.
The query form returns the status of the option.
Group
Trigger
Syntax
TRIGger:AUTOSync:DATARate{ON|OFF|0|1}
TRIGger:AUTOSync:DATARate ?
Arguments
ON enables the AutoSync data rate option.
OFF disables the AutoSync data rate option.
0 disables the AutoSync data rate option.
1 enables the AutoSync data rate option.
Examples
TRIGGER:AUTOSYNC:DATARATE ON sets the AutoSync data rate selection option
to ON, so that AutoSync will synchronize the data rate setting when it executes.
TRIGGER:AUTOSYNC:DATARATE ? might return
TRIGGER:AUTOSYNC:DATARATE 0, indicating that AutoSync is not set to
syncthedataratewhenitexecutes.
TRIGger:AUTOSync:DCRAtio
This command sets the AutoSync data-to-clock ratio selection option to ON
or OFF, so that when
AutoSync executes it will or will not synchronize the
data-to-clock ratio to the currently selected waveform.
The query form returns the status of the option.
Group
Trigger
2-302 DSA/CSA/TDS8X00/B Series Programmer Manual