EasyManua.ls Logo

Tektronix DSA8200 Series

Tektronix DSA8200 Series
431 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commands Listed in Alphabetical Order
Examples
TRIGGER:CLKRE
C:CRC:HALFRATE?
might return 1, indicating that the clock
recovery instrument is running at half the clock rate.
TRIGger:CLKRec:CRC:HWRev? (Query Only)
This query only command returns the hardware revision string of the connected
Clock Recovery instrument.
Group
Trigger
Syntax
TRIGger:CLKRec:CRC:HWRev?
Examples
TRIGGER:CLKREC:CRC:HWREV? might return :TRIGGER:CLKREC:CRC:HWREV
xxxx
(xxxx represents the hardware revision string).
NOTE. undened is returned if a Clock Recovery instrument is not present.
TRIGger:CLKRec:CRC:LOCKCount? (Query Only)
This query only command returns the number of times the connected Clock
Recovery instrument has relocked since the last reset.
Group
Trigger
Syntax
TRIGger:CLKRec:CRC:LOCKCo unt?
Examples
TRIGGER:CLKREC:CRC:LOCKCO UNT? might return
:TRIGGER:CLKREC:CRC:LOCKC OUNT 1 indicating t hat the Clock Recovery
instrument has locked once.
NOTE. 0 is returned if a Clock Recovery instrument is not present.
TRIGger:CLKRec:CRC:LOCKFp (No Query Form)
Locks or unlocks the clock recovery instrument front panel controls.
Group
Trigger
2-316 DSA/CSA/TDS8X00/B Series Programmer Manual

Related product manuals