Commands Listed in Alphabetical Order
Examples
TRIGGER:CLKRE
C:CRC:HALFRATE?
might return 1, indicating that the clock
recovery instrument is running at half the clock rate.
TRIGger:CLKRec:CRC:HWRev? (Query Only)
This query only command returns the hardware revision string of the connected
Clock Recovery instrument.
Group
Trigger
Syntax
TRIGger:CLKRec:CRC:HWRev?
Examples
TRIGGER:CLKREC:CRC:HWREV? might return :TRIGGER:CLKREC:CRC:HWREV
xxxx
(xxxx represents the hardware revision string).
NOTE. undened is returned if a Clock Recovery instrument is not present.
TRIGger:CLKRec:CRC:LOCKCount? (Query Only)
This query only command returns the number of times the connected Clock
Recovery instrument has relocked since the last reset.
Group
Trigger
Syntax
TRIGger:CLKRec:CRC:LOCKCo unt?
Examples
TRIGGER:CLKREC:CRC:LOCKCO UNT? might return
:TRIGGER:CLKREC:CRC:LOCKC OUNT 1 indicating t hat the Clock Recovery
instrument has locked once.
NOTE. 0 is returned if a Clock Recovery instrument is not present.
TRIGger:CLKRec:CRC:LOCKFp (No Query Form)
Locks or unlocks the clock recovery instrument front panel controls.
Group
Trigger
2-316 DSA/CSA/TDS8X00/B Series Programmer Manual

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA8200 Series and is the answer not in the manual?

Tektronix DSA8200 Series Specifications

General IconGeneral
BrandTektronix
ModelDSA8200 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals