Commands Listed in Alphabetical Order
Examples
TRIGGER:PSYNC
:TRIGPROBE:PLENGTH 128
This command deļ¬nes the pattern
length for the module connected to t he TRIGGER PROBE POWER connector as
128.
TRIGGER:PSYNC:TRIGPROBE:PL ENGTH? might return
TRIGGER:PSYNC:TRIGPROBE:PL ENGTH 128, indicating the current
pattern length is 128 for the module connectedtotheTRIGGERPROBEPOWER
connector.
TRIGger:SETLevel (N
o Query Form)
This command (no query form) sets the trigger level automatically to 50% of the
minimum/maximum value of the trigger input signal. This trigger level is effective
only when the trigg
er source is set to External Direct (see the TRIGger:SOUrce
command). The trigger level is the voltage threshold through which the trigger
source signal must pass to generate a trigger event. This command is equivalent
to clicking Set to 50% in the Level section of the Trigger Setup dialog box or
pressing the Trigger Set to 50% button on the front panel.
Group
Trigger
Syntax
TRIGger:SETLevel
Related Commands
TRIGger:SOUrce, TRIGger:LEVel
Examples
TRIGGER:SET
LEVEL
sets the trigger level to 50% of the applied signal.
TRIGger:SLOpe
This command sets or queries the trigger slope. The slope is applied only when
trigger source is set to External Direct (see the TRIGger:SOUrce command). (For
External Prescaler, Internal Clock, and Clock Recovery trigger sources, the slope
is set internally to rising.) Sending this command is equivalent to selecting the
rising or falling icon in the Slope section of the Trigger Setup dialog box.
Group
Trigger
Syntax
TRIGger:SLOpe {RISe | FALL}
TRIGger:SLOpe?
Related Commands
TRIGger:SOUrce
DSA/CSA/TDS8X00/B Series Programmer Manual 2-339