EasyManua.ls Logo

Tektronix DSA8200 Series

Tektronix DSA8200 Series
431 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commands Listed in Alphabetical Order
Related Commands
CH<x>:SCAle, C
URVe
Arguments
NR3 is the vertical scale factor in YUNits (usually volts) per division of the
incoming waveform.
Examples
WFMINPRE:YMULT 5.0E -3 sets the vertical scale factor of the incoming
waveform to 5 mV per division.
WFMINPRE:YMULT? might return WFMINPRE:YMULT 0.100000000000,
indicating that the vertical scale of the incoming waveform is set to 100 mV
per division.
WFMInpre:YOFf
This command sets or queries the vertical position of the incoming waveform
in divisions.
Group
Waveform Transfer
Syntax
WFMInpre:YOFf <NR3>
WFMInpre:YOFf?
Related Commands
CH<x>:POSition
Arguments
NR3 is the vertical position in divisions of the incoming waveform.
Examples
WFMINPRE:YOFF 1.0E+0 sets the vertical position to 1 division.
WFMINPRE:YOFF? might return WFMINPRE:YOFF -2 indicating that the vertical
position of the incoming waveform is -2 divisions.
WFMInpre:YSCALE
This command sets or queries the vertical unit scale factor of the incoming
waveform. The value in vertical units is calculated by the following formula:
Vertical Units = (Raw Data) * YSCALE + YZERO.
Group
Waveform Transfer
2-362 DSA/CSA/TDS8X00/B Series Programmer Manual

Related product manuals