Commands Listed in Alphabetical Order
Related Commands
CH<x>:SCAle, C
URVe
Arguments
NR3 is the vertical scale factor in YUNits (usually volts) per division of the
incoming waveform.
Examples
WFMINPRE:YMULT 5.0E -3 sets the vertical scale factor of the incoming
waveform to 5 mV per division.
WFMINPRE:YMULT? might return WFMINPRE:YMULT 0.100000000000,
indicating that the vertical scale of the incoming waveform is set to 100 mV
per division.
WFMInpre:YOFf
This command sets or queries the vertical position of the incoming waveform
in divisions.
Group
Waveform Transfer
Syntax
WFMInpre:YOFf <NR3>
WFMInpre:YOFf?
Related Commands
CH<x>:POSition
Arguments
NR3 is the vertical position in divisions of the incoming waveform.
Examples
WFMINPRE:YOFF 1.0E+0 sets the vertical position to 1 division.
WFMINPRE:YOFF? might return WFMINPRE:YOFF -2 indicating that the vertical
position of the incoming waveform is -2 divisions.
WFMInpre:YSCALE
This command sets or queries the vertical unit scale factor of the incoming
waveform. The value in vertical units is calculated by the following formula:
Vertical Units = (Raw Data) * YSCALE + YZERO.
Group
Waveform Transfer
2-362 DSA/CSA/TDS8X00/B Series Programmer Manual