Commands Listed in Alphabetical Order
Arguments
Qstring isaquo
ted s tring.
Examples
WFMINPRE:XUNIT "S" sets the horizontal units t
o seconds.
WFMINPRE:XUNIT? might return WFMINPRE:XUNIT "S", indicating that the
horizontal units of the incoming waveform are
seconds.
WFMInpre:XZEro
This command sets or queries the horizontal (x-axis) origin offset (time of first
point in this instrument) of the incoming w
aveform. The time of first point is the
time interval between the trigger point and the first point in the record (waveform).
Group
Waveform Transfer
Syntax
WFMInpre:XZEro <NR3>
WFMInpre:XZEro?
Related Commands
HORizontal:MAIn:TOFPoint?
Arguments
NR3 is the time interval between the trigger point and the first point in the
waveform record.
Examples
WFMINPRE:XZERO 20E-9 sets the horizontal origin offset (time of first point) to
20 ns for the incoming waveform. This is the time interval between the trigger
point and the fi rst point in the waveform record.
WFMINPRE:XZERO? might return WFMinpre:XZero 5 .000000000000E-8,
indicating that the time interval between the trigger point and the first point in the
incoming waveform record is 50 ns.
WFMInpre:YMUlt
This command sets or queries the vertical scale in v ertical units per d ivision of the
incoming waveform returned by the previous CURVe? command.
Group
Waveform Transfer
Syntax
WFMInpre:YMUlt <NR3>
WFMInpre:YMUlt?
DSA/CSA/TDS8X00/B Series Programmer Manual 2-361