Commands Listed in Alphabetical Order
will be returne
d as possible without exceeding the 60-character limit of the
<Message> and <Command> strings combined. The command string is right
justiļ¬ed.
Examples
ALLEV? might return the string :ALLEV 2225,"MEASUREM ENT ERROR, NO
WAVEFORM TO MEASURE;", 420,"QUERY UNTERMINATED;"
.
APPlication:ACTivate
(No Query Form)
This command (no query form) is used to launch the optional software
applications. The application must be installed and activated with a key code. (A
key code is not necess
ary if the application is within a free trial period).
Group
Miscellaneous
Syntax
APPlication:ACTivate {80SJNB}
Arguments
80SJNB is the optional 80SJNB Advanced Jitter, Noise, and BER Analysis
software.
80SJNB is a comprehensive jitter, noise, and bit error ratio (BER)
analysis application for serial data signal impairment characterization. The
80SJNB application requires approximately 10 seconds to launch.
Examples
APPLICATION:ACTIVATE 80SJNB launches the 80SJNB software application.
AUTOSet (No Query Form)
This command (no query form) causes the instrument to adjust its vertical,
horizontal, and trigger controls to provide a stable display of the selected
waveform. Sending this command is equivalent to pressing the front-panel
AUTOSET button.
NOTE. If the currentl
y selected waveform is a mask source, a mask autoset will
be performed.
Group
Miscellaneous
Syntax
AUTOSet EXECute
2-66 DSA/CSA/TDS8X00/B Series Programmer Manual