ication
Table 25: Delta Time Measurement Accuracy (cont.)
MDO V/Div So urce V
pp
Test Result High Limit
5mV 40mV 3.68s
100 mV 800 mV 2.74 s
500 mV 4 V 2.70 s
1V 4V 5.35s
Channel 3
1
MDO = 4 ns/Div, Source frequency = 240 MHz (does not apply to 100 and 200 MHz models)
MDO V/Div So urce V
pp
Test Result High Limit
100 mV 800 mV 233 ps
500 mV 4 V 233 ps
1 V 4 V 237 ps
MDO = 40 ns/Div, Source frequency = 24 MHz
MDO V/Div So urce V
pp
Test Result High Limit
5 mV 40 mV 435 ps
100 mV 800 mV 359 ps
500 mV 4 V 356 ps
1 V 4 V 583 ps
MDO = 400 ns/Div, Source frequency = 2.4 MHz
MDO V/Div So urce V
pp
Test Result High Limit
5mV 40mV 3.69ns
100 mV 800 mV 2.75 ns
500 mV 4 V 2.71 ns
1 V 4 V 5.36 ns
MDO = 4 μs/Div, Source frequency = 240 kHz
MDO V/Div So urce V
pp
Test Result High Limit
5mV 40mV 36.8ns
100 mV 800 mV 27.4 ns
500 mV 4 V 27.0 ns
1V 4V 53.5ns
MDO = 40 μs/Div, Source frequency = 24 kHz
MDO V/Div So urce V
pp
Test Result High Limit
5 mV 40 mV 368 ns
100 mV 800 mV 274 ns
500 mV 4 V 270 ns
54 MDO3000 Series Specications and Performance Verication