Performance Ver
ification
DVM Voltage Accuracy Tests ( AC)
Table 39: DVM Voltage Accuracy (AC)
Channel 1
Vertical Scale Input Signal
Low Limit Test Result High Limit
5mV 20mV
pp
at 1 kH
z
9.800 mV 10.200 mV
10 mV 50 mV
pp
at 1 kHz 24.5 m V 25.500 mV
100 mV 0.5 V
pp
at 1 kHz 245.000 mV 255.000 mV
200 mV 1 V
pp
at 1 k Hz 490.000 mV 510.000 mV
1V 5V
pp
at 1 kHz 2.450 mV 2.550 mV
Channel 2
Vertical Scale Input Signal
Low Limit Test Result High Limit
5mV 20mV
pp
a
t1kHz
9.800 mV 10.200 m
V
10 mV 50 mV
pp
at 1 kHz 24.5 m V 25.500 mV
100 mV 0.5 V
pp
at 1 kHz 245.000 mV 255.000 mV
200 mV 1 V
pp
at
1kHz
490.00
0mV
510.00
0mV
1V 5V
pp
at 1 kHz 2.450 mV 2.550 mV
Chann
el 3
1
Vertical Scale Input Signal
Low Limit Test Result High Limit
5mV 20mV
pp
at 1 kHz
9.80
0mV
10.2
00 mV
10 mV 50 mV
pp
at 1 kHz 24.5 m V 25.500 mV
100 mV 0.5 V
pp
at 1 kHz 245.000 mV 255.000 mV
200
mV
1V
pp
at 1 kHz
490
.000 mV
510
.000 mV
1V 5V
pp
at 1 kHz 2.450 mV 2.550 mV
Ch
annel 4
1
Vertical Scale Input Signal
Low Limit Test Result High Limit
5
mV
2
0mV
pp
at 1 kHz
9
.800 mV
1
0.200 mV
10 mV 50 mV
pp
at 1 kHz 24.5 m V 25.500 mV
100 mV 0.5 V
pp
at 1 kHz 245.000 mV 255.000 mV
200 mV 1 V
pp
at 1 k Hz 490.000 mV 510.000 mV
1V 5V
pp
at 1 kHz 2.450 mV 2.550 mV
1
Channels 3 and 4 are only on four-channel oscilloscopes.
MDO3000 Series S pecifications and Performance Verification 69