EasyManua.ls Logo

Tektronix MDO3000 Series

Tektronix MDO3000 Series
125 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Ver
ication
Random Noise, Sample Acqu isition Mode Tests
Table 24: Random Noise, Sample Acquisition Mode
Random Noise,
Sample Acquisition Mode
Bandwidth Sel
ection
Test result High limit
Full 4.50 mV
250 MHz 4.15 mV
Channel 1
20 MHz 4.10 mV
Full 4.50 mV
250 MHz 4.15 mV
Channel 2
20 MHz 4.10 mV
Full 4.50 mV
250 MHz 4.15 mV
Channel 3
1
20 MHz 4.10 mV
Full 4.50 mV
250 MHz 4.15 mV
For instruments with
1 GHz bandwidth
(includes MD
O310X
models as well as
MDO305X/303X/302
X/301X mode
ls with
1 GHz upgrade)
Channel 4
1
20 MHz 4.10 mV
Full 4.15 mV
250 MHz 4.15 mV
Channel 1
20 MHz 4.10 mV
Full 4.15 mV
250 MHz 4.15 mV
Channel 2
20 MHz 4.10 m
V
Full 4.15 mV
250 MHz 4.15 mV
Channel 3
1
20 M
Hz
4.1
0mV
Full 4.15 mV
250 MHz 4.15 mV
For instruments with
500 MHz bandwidth
(inclu
des MDO305X
models as well as
MDO303X/302X/301X
model
s with 500 MHz
upgrade)
Channel 4
1
20
MHz
4.
10 mV
50 MDO3000 Series Specications and Performance Verication

Table of Contents

Other manuals for Tektronix MDO3000 Series

Related product manuals