Performance Ver
ication
Table 24: Random Noise, Sample Acquisition Mode (cont.)
Random Noise, Sa mp le Acquisition Mode Bandwidth Selection Test result High limit
Full 4.15 mV
250 MHz 4.15 mV
Channel 1
20 MHz 4.10 mV
Full 4.15 mV
250 MHz 4.15 mV
Channel 2
20 MHz 4.10 mV
Full 4.15 mV
250 MHz 4.15 mV
Channel 3
1
20 MHz 4.10 mV
Full 4.15 mV
250 MHz 4.15 mV
For instruments with
350 MHz bandwidth
(includes MDO303X
models as well as
MDO302X/301X
models with 350 MHz
upgrade)
Channel 4
1
20 MHz 4.10 mV
Full 4.15 mV
Channel 1
20 MHz 4.10 mV
Full 4.15 mV
Channel 2
20 MHz 4.10 mV
Full 4.15 mV
Channel 3
20 MHz 4.10 mV
Full 4.15 mV
For instruments with
200 MHz bandwidth
(MDO302X models
as well as MDO301X
models with 200 MHz
upgrade)
Channel 4
20 MHz 4.10 mV
Full 4.15 mV
Channel 1
20 MHz 4.10 mV
Full 4.15 mV
Channel 2
20 MHz 4.10 mV
Full 4.15 mV
Channel 3
1
20 MHz 4.10 mV
Full 4.15 mV
For instruments with
100 MHz bandwidth
(MDO301X models)
Channel 4
1
20 MHz 4.10 mV
1
Channels 3 and 4 are only on four-channel oscilloscopes.
MDO3000 Series S pecications and Performance Verication 51